29
4.
Hold the assembly in a vertical, upright position and tap against it gently until the filter
falls out.
5.
Insert the new measurement filter into the housing. The arrow on the filter indicates
the direction of installation. Insert the filter into the housing with the arrow pointing
down.
6.
1.5 mm
Tighten the securing screw of the measurement filter until it is in contact with the filter.
Make sure that the filter is held in position.
7.
Screw the detector module into the detector housing.