Maintenance
OUSAF44
28
Hauser
In addition, you must reset the counter for filter replacement in the case of the CM44P:
CAL
/
Photometer
/
Filter change
/
Reset
8.5
Replacing the measurement filter
Versions with Easycal:
If the Easycal detector module is sent in for recertification, a diagnostic scan of the
measurement filter is performed and the measurement filter is replaced if necessary.
Do not replace the filter yourself.
Versions with a standard detector
1.
Remove the detector housing from the flow assembly.
2.
1"
Unscrew the detector module in a counterclockwise direction to remove it from the
detector housing. If it is stuck, place a 1" (25 mm) open-ended wrench on the base plate
of the cable socket to unscrew the assembly.
3.
1.5 mm
Release the fixing screw.