Maintenance
OUSAF44
30
Hauser
8.
Mount the detector housing back onto the flow assembly.
You must then recalibrate/readjust the measuring system.
In addition, you must reset the counter for filter replacement in the case of the CM44P:
CAL
/
Photometer
/
Filter change
/
Reset
8.6
Replacing the sensor window and seal
Operating Instructions for Flowcell OUA260, BA01600C
Operating Instructions for CUA261, BA01652C
If you have mounted the sensor in a VARIVENT flow assembly using the CUA261
adapter, refer to the Operating Instructions of the adapter for information on removal
and on replacing the optical windows.
Removing optical windows and seals
Always replace windows with windows of the same type in order to maintain the path length.
The following applies for OUA260:
The flow assembly must be removed from the process line in order to replace the windows
and seals.
1. In the case of OUA260 only:
Stop the flow in the process pipe and remove the assembly from the
dry
process line.
2. Remove the lamp and detector housing from the assembly.