6 Start-up
Hauser
51
6.5.10
Set up 2 – Data Log
The data logger records two freely selectable parameters with their date and time. You
can start it using the measuring menus:
Use the arrow keys to scroll through the measuring menus until to you reach the Record
mode of the data logger. Pressing the "Enter" key brings you to the Scroll mode of the
data logger. Here you can open the saved measured values with their date and time.
To enter the menu, proceed as follows:
⇒
⇒
PARAM
pH
7.00
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next
E
Hold
pH
7.00
Param
Edit
Next
Set up 2
E
Hold
Limit switch
Data log
Contr. quick adj.
Controller settings
Check systems
CODE
CHOICE
(default = bold)
INFO
K1
Measuring interval
Data logger1
Data logger2
DataLog View 1
DataLog View 2
Data logger settings
Using the data logger you can record
• one parameter with 500 sequential measuring points or
• two parameters each with 500 sequential measuring
points.
Measuring interval:
KA1
00005s
(2 ... 36000 s)
Enter measuring interval
Enter the time interval after which the next measured
value is recorded in the data logger.
Data logger 1 (or 2):
KB1 / KC1
Meas-
ured
value:
Function:
pH/redox K1
on
Selection
Set the measured variable for recording (pH/redox ,
temp.) and the activate using the "on" function.
!
Note!
The data logger starts recording the measured value
when you return to operation.
KB2 / KC2
Min:
Max:
12.00pH/-500mV
12.00pH/500mV
(-2...16pH/
-1500...1500mV)
Set recording range
Values outside the defined range are not recorded.
DataLog View 1 (or 2)
KD1
View of recorded data
7.54
12 15 35
01 04 2001
pH 7.54
[ ¯]
Select
Para
DataLog View 1