6 Start-up
Mycom S CPM 153
60
Hauser
To enter the menu, proceed as follows:
⇒
⇒
PARAM
pH
7.00
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next
E
Hold
pH
7.00
Param
Edit
Next
Set up 2
E
Hold
Check systems
Limit switch
Data log
Contr. quick adj.
Controller settings
CODE
CHOICE
(default = bold)
INFO
M1
off
on
Selection of Controller settings
!
Note!
You must activate the controller settings
after
you have
configured the controller in this menu branch.
M2
Batch one-sided
Inline one-sided
Batch two-sided
Inline two-sided
Select the process type,
which describes your process.
One-sided:
control using either acid
or
alkali
Two-sided:
control using either acid
and
alkali.
You can only select this function if you have defined two
controllers (in the "Contacts" menu and/or via the cur-
rent output).
M3
Actuators
Sensors
Feedback
Curve
Select external hardware
For correct operation, you must completely configure
these four submenus.
Actuators
: here, you can select and configure the
methods which the controller uses to output the set
values.
Sensors:
here you configure the look-ahead pH measu-
rement or switch channels (only with two-circuit).
Feedback:
here you configure the position feedback of
an actuator drive (only with the selection of three PS
and position feedback = on; see Fields 162, 165 / 170,
165)
Curve:
here, you enter the controller parameters (neu-
tral zone, set point, etc.). With this selection, you can
also reach the "active measuring menu" (see Field 196).
Actuators:
With selection "one-sided" in field M1:
MA1
Acid
Alkali
Dosing
Select the medium to be dosed to the process.
MA2
Pulse length
Pulse-frequency
3-point step controller
Current output
Select control type