6 Start-up
Mycom S CPM 153
42
Hauser
Medium compensation (only for pH):
GB1
Select table
Create tables
Reference temperature
Selection
Enter / activate customer-specific temperature compen-
sation tables.
Select table
= select for activation
Select table:
GBA1
Medium 1
Medium 2
Medium 3
off
Select medium for measuring circuit 1
Select a medium for measuring circuit 1.
off
= no medium compensation
GBA2
Medium 1
Medium 2
Medium 3
off
Select medium for measuring circuit 2
(only two circuit instruments)
Select a medium for measuring circuit 2.
off
= no medium compensation
Create tables:
GBB1
Medium 1
Medium 2
Medium 3
Select medium
Medium compensation curves can be entered
as a table for three different media.
GBB2
02
(2 ... 10)
Entry of the number of support points (value pairs)
Value pair: pH/redox and temperature
GBB3
°C
020.0°C
025.0°C
pH
02.00
04.00
Value pair entry
Enter pH/redox and temperature (number of required
value pairs = number of support points desired in Field
GBB2).
GBB4
OK
Delete element(s)
Selection:
Are the value pairs OK or do you want to delete
elements?
GBB5
°C
020.0°C
025.0°C
pH
02.00
04.00
Delete:
Select the row to be deleted, delete with
T
and
confirm by pressing "E".
GBB6
Valid table
Note in display:
The table is active after confirmation by pressing "E". Can-
cel by pressing "PARAM".
Reference temperature:
GBC1
For laboratory measu-
rement:
25.0°C
(0 ... 100°C)
Enter reference temperature
to which the medium temperature shall be
compensated. Enter the temperature at the pH r
eference value of the process is defined (e.g. the
ambient temperature in the laboratory).
CODE
CHOICE
(default = bold)
INFO