6 Start-up
Mycom S CPM 153
74
Hauser
6.5.16
Manual operation
To enter the menu, proceed as follows:
⇒
⇒
PARAM
pH
7.00
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next
E
Hold
pH
7.00
Param
Edit
Next
Manual operation
E
Hold
HOLD off
HOLD on
CODE
CHOICE
(Default = bold)
INFO
R1
ChemoClean
Hold
Select manual operation
!
Note!
• Leave the manual operation menu by pressing
"PARAM", "DIAG" or "MEAS".
• The settings are only active in this menu. Nothing is
saved when you leave.
R2
!!!Caution!!
You are now leaving manual
operation.
If you leave the manual operation:
Note in display
Confirm with "Enter": Leave the manual operation.
Abortion with "PARAM": Remaining in manual operation
mode.
ChemoClean:
RB1
Weekly progr.
Clean trigger
Ext. Control
off
off
off
Note in display (no entry):
System status
RB2
Abort
Start
ChemoClean cleaning
Start / Abort.
Here, each external programme start is suppressed.
!
Note!
Leave this item by pressing "PARAM".
HOLD:
RC1
HOLD off
HOLD on
Manual operation
Activate / deactivate Hold
The "HOLD" function freezes the current outputs as
soon cleaning/calibration is undertaken.
!
Note!
If the controller function lies on current output 2,
it follows the instructions of the defined "controller hold"
(s. page 45).