![Endress+Hauser Mycom S CPM 153 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/endress-hauser/mycom-s-cpm-153/mycom-s-cpm-153_operating-instructions-manual_2408552047.webp)
Mycom S CPM 153
6 Start-up
Hauser
47
JB4
°C:
000.0
005.0
...
pH:
04.00
04.05
...
Value pair entry
Enter temperature and pH/Redox (number of required
value pairs = number of support points desired in
field JB3).
JB5
OK
Delete element(s)
Selection:
Are the value pairs OK or do you want
to delete any of them?
JB6
°C:
000.0
005.0
...
pH:
04.00
04.05
...
Delete:
Select the row to be deleted, delete it with
T
and
confirm this with "E".
JB7
Valid table
Note in display:
The table is active after confirmation by pressing "E".
Cancel by pressing "PARAM".
Cal. settings:
JC1
MTC
ATC 1
ATC 2
Select the temperature compensation for the
calibration
ATC = automatic temp. comp.
MTC = manual temp. comp.
!
Note!
The setting is only active during calibration. In meas-
urement mode, the setting selected in GAA1 is valid.
JC2
5.00 mV/pH
(5.00 ... 57.00 mV/pH)
Entry of difference to slope for alarm function
If the entered slope difference is exceeded, an alarm
(error no. 032 / E035) can be triggered (error activation
in field H5).
JC3
pH 1.30
(0.05 ... 2.00 pH)
Entry of pH value zero point deviation
for the alarm function
If the zero point deviates from the reference zero point
by the value entered here, an alarm (Error no. 033) can
be triggered (error activation in field H5).
JC6
off
on
SCC (Sensor Condition Check)
This function monitors the electrode status or the
degree of electrode ageing. Possible status mes-
sages: "Electrode OK", "Low wear" or "Replace elec-
trode". The electrode status is updated after each cali-
bration. When the "Replace electrode" message
appears, an error message may be displayed.
!
Note!
This function is only available for glass electrodes. If
you are using a glass electrode and an IsFET sensor,
you can use the SCC function without restriction.
However, the SCC function only monitors the glass
electrodes.
CODE
CHOICE
(default = bold)
INFO