8 Troubleshooting
Hauser
97
8.4.2
Coding
Current outputs active or passive:
For instrument versions CLM 153 xxA/Bxx (2 current outputs) and CLM 153 xxC/Dxx
(2 current outputs with HART) the current outputs can be operated as active or passive.
Jumpers on the controller module M3CH allow recoding.
For non-Ex instruments, these modules may be recoded to active outputs.
#
Warning!
Ex instruments must
not
be recoded. Doing so will cause loss of intrinsic safety!
.
8.5
Replacing the device fuses
For non-Ex devices
#
Warning!
Danger to persons. Before replacing the fuse, make sure the device is voltage-free.
•
Position of the fuse holder: "A" in fig. 28
•
Use only a 5 x 20 mm fine-wire fuse with 3.15 mA, semi time-lag fuse. All other fuses
are not permitted.
"
Caution!
If the fuse should fail again, have the device checked.
8.6
Disposal
The Mycom S CLM 153 is a transmitter which contains electronic components and
PCBs and therefore must be disposed of as electronic refuse. Please keep to the local
regulations.
Fig. 29:
Coding for current outputs
(Interior view of the housing upper section)
Fig. 30:
Active/passive coding of the current
outputs
1
2
C07-CPM153xx-09-06-00-xx-001.EPS
1
2
1
1
1
1
2
1
Jumper setting
passive output
Jumper setting
active output
C07-CPM153xx-09-06-00-en-002.EPS