6 Commissioning
Mycom S CLM 153
28
EHauser
T17
NAMUR
Relay 1:
Relay 2:
off
free
free
Contact functions
Depending on the equipment available, you
can assign the function of up to 5 relays here.
When you switch on the status message acc.
to NAMUR NA64, relays 1 and 2 are assigned
and are not available for another function,
(compare page 14).
Selection: Free / Controller / LC / CCW / CCC
Controller:
Controller control using relay
LC
: Limit contactor function
CCW
: ChemoClean water. Water supply for
the ChemoClean function.
CCC
: ChemoClean Cleaner. Cleaner supply
for the ChemoClean function.
(Together, CCC and CCW form the
"ChemoClean" function. You can find
information on ChemoClean on page 70)
!
Note!
If you want to use the USP function, select the
limit contactor function for a relay and
configure this in the limit value menu for USP
(p. 68).
T18
Measured value CH1
Measured value CH2
Temperature CH1
Temperature CH2
Related
Select the measured value
which should be output at the current
output 1.
Selection possibilities related to the instrument
variant and the selected output.
Measured value 1/2
: Selection of the primary
measured value which you selected,
(conduct., conc., resist.)
Temperature 1/2
: Selection of the temperature
to be output to the current output.
Related (only for related circuits):
The
characteristic value from Field T5 is output at
the current output.
Continuous controller (only at current
output 2!)
: Controller output
T19 Measured
value
CH1
Measured value CH2
Temperature CH1
Temperature CH2
Related
Continuous controller
(only
at current output 2)
Select measured value
which should be output at the current
output 2.
Selection possibilities see above,
Continuous controller (only at current output
2!)
: The controller actuating variable is output
via the current output (see also Controller
menu page 54).
!
Note!
Danger of data loss.
If you change the
assignment for the current output from
"continuous controller" to a different function
after
you have configured the controllers, the
entire
controller configuration (s. page 54) is
reset to the default values.
CODE
DISPLAY
CHOICE
(default = bold)
INFO
Hold
Param
E
3
mS/cm
.52
Relay funct.
Next
Acc.Namur
off
Select
Relais 1
free
Relais 2
free
Hold
Param
Edit
E
3
mS/cm
.52
Current out 1
Temp. CH1
Temp. CH2