6 Commissioning
Mycom S CLM 153
80
EHauser
To enter the menu, proceed as follows:
⇒
⇒
Cal
Next
Selection
E
Hold
Cell constant
Adaption factor
Airset
3.52 mS/cm
Edit
CODE
CHOICE
(default = bold)
INFO
C1
Sensor 1
Sensor 2
End calibration
Selection for calibration
(only two circuit)
Select sensor 1
or
2, and then run through calibration
for each individual sensor.
C2
Cell constant
Airset
Adaption factor
Selection
Cell constant
: Calculate the cell constant of the sensor.
Airset
(only inductive): The sensor must be calibrated
at air and in a dry state.
Adaption factor
(only inductive): Sensor calibration with
compensation for the wall effect.
Cell constant:
Immerse the sensor (inductive or conductive)
into the calibration solution.
!
Note!
Immerse the sensor in such a way that there is
a minimum distance of 15 mm to the vessel
wall, so that the installation factor does not play
a role.
CB1
025.0 °C
(–35.0 ... 250 °C)
Entry of MTC temperature
(only MTC)
CB2
2.10 % / K
(0.00 ... 20.00% / K)
Alpha value
Enter the
α
value of the calibration solution. This value
is specified for all E+ H calibration solutions.
CB3
Current measured value
(0.0 ... 9999 mS/cm)
Current measured value
Enter the correct conductivity value of the calibration
solution.
It makes sense to use a calibration solution which is at
>40 % of the measuring range of the sensor used. The
display is always in mS/cm.
CB4
5.9 cm-1
(0.1 ... 9.99 cm-1)
Cell constant
The calculated cell constant is displayed and applied
to Field A5 (s. page 31).
CB5
ok
Calibration status
CB6
Accept
Reject
Recalibrate
End of calibration
Accept
: Pressing "E" accepts the new calibration data.
Reject
: The data are not accepted, a recalibration will
not be started.
Recalibrate
: The data are rejected and a new
calibration will be started.
Proceed with
F
C07-CLM153xx-16-06-00-xx-003.EPS
~
~ ~
~
~
~
~