6 Commissioning
Mycom S CLM 153
50
EHauser
6.4.12
Set up 2 – Data log
The data logger records two freely selectable parameters with their date and time. You
can start it using the measuring menus:
Use the arrow keys to scroll through the measuring menus until to you reach the Record
mode of the data logger. Pressing the "Enter" key brings you to the Scroll mode of the
data logger. Here you can open the saved measured values with their date and time.
To enter the menu, proceed as follows:
⇒
⇒
PARAM
Param
Settings
Set up 1
Set up 2
Manual operation
First start up
Edit
Next E
Hold
3.52 mS/cm
Param
Edit
Next
Set up 2
E
Hold
Limit switch
Controller settings
Data log
Contr. quick adj.
Check
3.52 mS/cm
CODE
CHOICE
(default = bold)
INFO
K1
Measuring interval
Data logger 1
Data logger 2
View 1
View 2
Data logger settings
Using the data logger you can
• Record a parameter with 500 sequential measuring
points or
• two parameters each with 250 sequential measuring
points.
View 1/2
: You can view the data recorded in the data
logger.
Measuring interval:
KA1
00005s
(2 ... 36000s)
Enter measuring interval
Enter the time interval after which the next measured
value is recorded in the data logger.
Data logger 1 (or 2):
KB1 / KC1
Measured
value:
Function:
Measured
value input 1
On
Selection
Set the measured value for recording (measured value
input 1, measured value input 2, temp. input 1, temp.
input 2, characteristic value) and then activate by
means of the "on" function.
!
Note!
The data logger starts recording the measured value
when you return to operation.
KB2 / KC2
Min:
Max:
0,00
2000,00
Set recording range
Values outside the defined range are not recorded.
View 1 (or 2):
KD1 / KE1
View of the recorded data
Using the date and time, you can call up data recorded
in the past.
0.00
3.52
12 15 35
01 04 2001
10
PS
mS/cm 1
Measure
Select