Mycom S CLM 153
10 Technical data
Hauser
103
Controller output
Relay contacts
Galvanic isolation
At the same potential are:
•
Current output 1 and power supply
•
Current output 2 and resistance input.
The remaining circuits are galvanically isolated from each other.
Electrical
connection
10.3
Performance characteristics
Measured value resolution
Conductivity:
0.001 µS/cm
Temperature: 0.1
K
Measured error
1
display
Conductivity, resistance, concentration:
±0.5% ±2 digits of measured value
Temperature:
< 0.5 K
Measured error
1
current
output
additionally max. 0.2% of current range maximum
Repeatability
1
Conductivity, resistance, concentration:
±0.2 % ±2 digits of measured value
Temperature:
max. 0.1% of measuring range
1
: acc. to IEC 746-1, under nominal operating conditions
Function (selectable):
Pulse-length controller (PWM)
Pulse-frequency controller (PFM)
Three-point step controller (3-PS)
Analogue (via current output)
Switch behaviour
P / PI / PID
Control gain K
R
0.01 … 20.00
Integral action time T
n
0.0 … 999.9 min
Derivative action time T
v
0.0 … 999.9 min
With the maximum settable frequency in PFM
120 min
-1
With the maximum settable period length in PWM
1 ... 999.9 s
With PWM minimum switch-on period
0.4 s
The NC/NO contact type can be set by software.
Switching voltage
max. 250 V AC / 125 V DC
Switching current
max. 3 A
Switching power
max. 750 VA
Lifespan
≥
5 million switching cycles
Power supply for CLM 153-xxxx
0
xxxx
100 ... 230 V AC +10/–15%
Frequency
47 ... 64 Hz
Power supply for CLM 153-xxxx
8
xxxx
24 V AC/DC +20/–15%
Power consumption
max. 10 VA
Separation voltage between galvanically isolated
circuits
276 V
rms
Terminals, max. cable cross-sectional area
3 x 2.5 mm²