Mycom S CLM 153
6 Commissioning
Hauser
31
A3
Mode:
Unit
Medium
Comb. unit
Cd.
auto
%
Select operating mode
(only related circuits)
If the operating mode changes, the user settings are
automatically reset. The settings you make here apply to
both measuring circuits.
Mode
: Cd. (conductivity, difference calculation), resistance,
concentration
Unit
(with Cond. operating mode, difference calculation):
auto, µS/cm, mS/cm, S/cm, µS/m, mS/m, S/m
Unit
(with resistance operating mode, difference calculation):
auto, k
Ω•
cm, M
Ω•
cm, k
Ω•
m, M
Ω•
m.
Unit
(with concentration operating mode, difference
calculation): ppm, mg/l, %, TDS.
The optimum unit is automatically selected with "auto".
Medium
(with "concentration" operating mode): NaOH,
HNO3, H3PO4, H2SO4, Table 1 to 4
Rel. unit (
ratio calculation)
:
Unit of the characteristic value
defined in the previous field. Options: none, %
A4
Measuring circuit 1
Measuring circuit 2
Select measuring circuit
Measuring channel 1 (or 2):
AA1
Mode:
Unit/
medium
Cond.
auto
Select operating mode
(not for related circuits)
If the operating mode changes, the user settings are
automatically reset.
Operating mode
: Cond. (conductivity), resistance,
concentration
Unit
(with cond. / resistance operating mode): auto, mS/cm,
µS/cm / auto, k
Ω
, M
Ω
.
Unit
(with concentration operating mode): ppm, mg/l, %,
TDS.
The optimum unit is automatically selected with "auto".
Medium
(with "concentration" operating mode): NaOH,
HNO3, H3PO4, H2SO4, Table 1 to 4
AA2
Cell
const.:
Cable
resist.
Adaption
factor
00.000cm-
1
00.00
Ω
Select measuring channel 1 or 2
Cell constant
: The exact cell constant is provided in the
sensor quality certificate.
Cable resistance
(for conductive): Enter the cable
resistance.
Adaption factor
(for inductive): Enter the adaption factor
here.
AA3
Measured
value:
Temperatur
e:
01 s
01 s
(01 ... 30 s)
Set measured value attenuation
The mean value over the set time is displayed.
CODE
CHOICE
(default = bold)
INFO