Mycom S CLM 153
6 Commissioning
Hauser
81
Airset (only inductive):
Remove sensor from the liquid and dry
thoroughly
.
!
Note!
As long as the sensor is out of the medium
during Airset, an error code referring to
measuring range violation (E055/E056) can be
ignored.
CA1
0.0µS/cm
Current measured value
Start calibration residual coupling using CAL key.
CA2
42µS/cm
(–80 ...80 µS/cm)
Residual coupling
The residual coupling of the system (sensor and
transmitter) is displayed.
CA3
ok
Calibration status
CA4
Accept
Reject
Recalibrate
End of calibration
Accept
: Pressing "E" accepts the new calibration data.
Reject
: The data are not accepted, a recalibration will
not be started.
Recalibrate
: The data are rejected and a new
calibration will be started.
Proceed with
F
Adaption factor (only inductive):
The sensor remains in the place of application.
CC1
025.0 °C
(–35.0 ... 250°C)
Entry of MTC temperature
(only MTC)
CC2
2.10 % / K
(0.00 ... 20.00% / K)
Alpha value
Enter the
α
value of the calibration solution. This value
is specified for all E +H calibration solutions.
CC3
Current measured value
(0.0 ... 9999 mS/cm)
Current measured value
Enter the correct conductivity value of the calibration
solution.
It makes sense to use a calibration solution which is at
>40 % of the measuring range of the sensor used. The
display is always in mS/cm.
CC4
1
(0.10 ... 5.00)
Installation factor
The calculated installation factor is displayed and
applied to Field A5 (s. page 31).
CODE
CHOICE
(default = bold)
INFO
C07-CLM153xx-16-06-00-xx-001.EPS
~
~ ~
~
~
~
~
~
~
~ ~
~
~
~
~
C07-CLM153xx-16-06-00-xx-002.EPS