5.3
Current outputs
The function group CURRENT OUTPUT is
used to configure the individual outputs.
Either a linear (O2 (1)) or, in conjunction with
the S version, a user-defined current output
characteristic (O2 (3)) can be entered.
Furthermore, a current output value can be
simulated to check the current outputs (O2 (2)).
Coding
Field
Selection or range
Factory setting
Display
Info
B3
Enter temperature
coefficient
α
2.10 %/K
0.00 ... 20.00 %/K
Only if B2 = lin.
Any table defined is not active in
this case.
B4
Enter process
temperature
25.0 °C
–35.0 ... 250.0 °C
Only if B1 = fixed.
Editing is only possible in °C.
B5
Temperature
sensor calibration
actual value display
–35.0 ... 250.0 °C
This entry is used to calibrate the
temperature sensor to an
external measurement.
Omitted if B1 = fixed.
B6
Temperature
difference is
displayed
0.0 °C
–5.0 ... 5.0 °C
The difference between the
temperatures measured and
entered is displayed.
Omitted if B1 = fixed.
Factory settings are printed in
bold
face;
base version does not include functions in italics.
Coding
Field
Selection or range
Factory setting
Display
Info
O
Function group
CURRENT
OUTPUT
Initial display in function group
CURRENT OUTPUT.
O1
Selection of
current output
Out1
Out2
A different characteristic can be
selected for each output.
O2 (1)
Enter linear
characteristic
lin
=
linear
(1)
sim = simulation (2)
table (3)
The slope of the characteristic
may be positive or negative.
See O2 (2), O2 (3) for the
simulation and table
characteristics.
B3
alphaVal
2.10
%/K
B4
ProcTemp.
25.0
°C
B5
RealTemp.
0.0
°C
B6
TempOffs.
0.0
°C
O
OUTPUT
O1
Sel.Out
Out1
O2
Sel.Type
lin
Instrument configuration
Liquisys M CLM 223 / 253
26
Hauser