Coding
Field
Selection or range
Factory setting
Display
Info
F3
Enter alarm delay
0 s
0 ... 2000 s (min)
Depending on the unit selected
in F2, the alarm delay is entered
in s or min.
F4
Selection of error
current
22 mA
2.4 mA
This selection must be made
even if all errors are suppressed
in F5.
F5
Selection of error
1
1 ... 255
This is where the errors are
selected that are to trigger an
alarm signal. The errors are
selected via the error number.
Please refer to the table in
chapter 7 for the error numbers.
The factory settings remain in
effect for all errors not edited.
F6
Set alarm contact
to be effective for
selected error
yes
no
If set to “no”, all the other alarm
settings (e.g. alarm delay) are
also deactivated. The settings
themselves are retained.
This setting
only
applies to the
error selected in F5. Factory
setting is
no
starting with E080!
F7
Set error current
to be effective for
selected error
no
yes
The error current selected in
F4 becomes effective or is
suppressed in case of error.
This setting
only
applies to the
error selected in F5.
F8
Automatic start of
cleaning function?
no
yes
This field does not exist for some
errors; see chapter 7.1.
F9
Return to menu or
select next error
next
=
next error
<---R
If next is selected, the display
returns to F5.
If <---R is selected, the display
returns to F.
Factory settings are printed in
bold
face;
base version does not include functions in italics.
F3
Err.Delay
0
s
22mA
F4
Err.Curr
F5
Sel.Error
1
yes
F6
Rel.Assg
no
F7
Curr.Assg
no
F8
CleanTrig
F9
Select
next
LM253E5A.CHP
Liquisys M CLM 223 / 253
Instrument configuration
Hauser
29