Coding
Field
Selection or range
Factory setting
Display
Info
S3
Manual hold
off
on
S4
Enter hold dwell
period
10 s
0 ... 999 s
S5
Enter SW upgrade
release code
(plus package)
0000
0000 ... 9999
Entry of an incorrect code returns
you to the measurement menu.
The number is edited with the
PLUS or MINUS key and
confirmed with the ENTER key.
S6
Enter SW upgrade
release code for
Chemoclean
0000
0000 ... 9999
Entry of an incorrect code returns
you to the measurement menu.
The number is edited with the
PLUS or MINUS key and
confirmed with the ENTER key.
S7
Order number is
displayed
The order code is changed
automatically to reflect an
upgrade.
S8
Serial number is
displayed
S9
Reset of
instrument
(restore default
values)
no
Sens = sensor data
Facty = factory settings
Facty = All data are erased and
reset to the factory
settings!
Sens = The sensor data are
erased.
S10
Perform
instrument test
no
Displ = display test
Factory settings are printed in
bold
face;
base version does not include functions in italics.
off
S3
Man. HOLD
10
S4
Cont.Time
s
0000
S5
PlusCode
0000
S6
CleanCode
order
S7
CD0005
SerNo
S8
12345678
no
S9
S.Default
no
S10
Test
LM253E5D.CHP
Liquisys M CLM 223 / 253
Instrument configuration
Hauser
47