5.8
Service
Coding
Field
Selection or range
Factory setting
Display
Info
K5
Selection of triplet
1
1 ... number of
triplets in K4
Any triplet can be edited.
K6
Enter
uncompensated
conductivity value
0.0 µS/cm
0.0 ... 9999 mS/cm
K7
Enter
concentration
value for K6
0.00 %
0.00 ... 99.99 %
K8
Enter temperature
value for K6
0.0 °C
–35.0 ... 250.0 °C
K9
Enter whether or
not the table
status is okay
yes
no
Return to K2.
Factory settings are printed in
bold
face;
base version does not include functions in italics.
Coding
Field
Selection or range
Factory setting
Display
Info
S
Function group
SERVICE
S1
Selection of
language
ENG
=
English
GER = German
FRA = French
ITA
= Italian
NEL = Dutch
ESP = Spanish
This field must be configured
once during start-up. Then you
can exit S1 and continue.
S2
Hold configuration
S+C
=
during setup
+ calibration
CAL = during
calibration
Setup = during setup
none = no hold
S = setup,
C = calibration.
K5
Sel.Elem.
1
K6
conduct.
0.0
µS/cm
K7
concentr.
0.0
%
K8
Temp.val.
0.0
°C
K9
Status ok
yes
S
SERVICE
ENG
S1
Language
S+C
S2
Auto HOLD
Instrument configuration