Technical data
Liquistation CSF28
68
Hauser
15.6 Environment
Ambient temperature
range
With cooling module:
-20 to 40 °C (0 to 104 °F)
Storage temperature
–20 to 60 °C (–4 to 140 °F)
Electrical safety
In accordance with EN 61010-1, protection class I, environment ≤ 2000 m (6500 ft) above
MSL. The device is designed for pollution degree 2.
Relative humidity
10 to 95%, not condensing
Degree of protection
• Front dosing compartment: IP 54
• Rear dosing compartment: IP 33
• Front panel with display (internal): IP 65
• Sample compartment: IP 54
Electromagnetic
compatibility (EMC)
Interference emission and interference immunity as per EN 61326-1:2013, Class A for
Industry
15.7 Process
Medium temperature range
2 to 50 °C (36 to 122 °F)
Process pressure range
Unpressurized, open channel (unpressurized sampling)
Medium properties
Vacuum pump
Sample media has to be free of abrasive substances.
Peristaltic pump
Sample media has to be free of abrasive substances.
Pay attention to the material compatibility of the wetted parts.
Process connection
•
Vacuum pump:
Intake hose ID 10 mm (3/8 in) and 13 mm (1/2 in)
•
Peristaltic pump:
Intake hose ID 10 mm (3/8 in)
15.8 Mechanical construction
Dimensions
See the "Installation" section → 12
Weight
Sampler version
Weight
Plastic version with refrigeration
101 kg (223 lbs)