Liquistation CSF28
Mounting
Hauser
13
Mounting conditions
Avoid siphoning effects in the suction line.
Do not route the suction line with an upward gradient to the sampling point.
Note the following when erecting the device:
• Erect the device on a level surface.
• Connect the device securely to the surface at the fastening points.
• Protect the device against additional heating (e.g. heaters or direct sunlight).
• Protect the device against mechanical vibrations.
• Protect the device against strong magnetic fields.
• Make sure air can circulate freely at the side panels of the cabinet. Do not mount the
device directly against a wall. Allow at least 150 mm (5.9 in.) from the wall to the left
and right.
• Do not erect the device directly above the inlet channel of a wastewater treatment plant.
5.1.3
Mechanical connection
Foundation plan
302 (11.9)
195.4 (7.69)
C
C
D
B
A
A
A
A
137.4 (5.41)
355 (14.0)
61 (2.4)
605 (23.8)
730 (28.7)
786 (30.9)
214 (8.43)
255 (10.0)
311 (12.2)
305 (12.0)
33.6 (1.32)
A0024406
5
Foundation plan. Unit of measurement mm (in)
A
Fasteners (4 x M10)
B
Cable inlet
C
Outlet for condensate and overflow > DN 50
D
Sample supply from below > DN 80
--- Dimensions of Liquistation