Maintenance
Liquistation CSF28
56
Hauser
1
2
6
3
4
5
A0013896
22
Vacuum pump
1
Outflow hose
2
Dosing glass
3
Dosing glass cover
4
Air hose connection
5
Lock for intake hose
6
Thread adapter nut for intake hose
1. Check the sample volume set in the
System
/
Sampling
menu. The volume can only
be changed in the Commissioning Wizard.
2. Slacken the thread adapter nut on the intake hose (item 6).
3. Turn the intake hose at the lock (item 5) to the "open" position and pull the hose
upwards to disconnect it.
4. Release the air hose (item 4) and remove the dosing glass (item 2) along with the
outflow hose (item 1) from the front.
5. Open the bayonet lock (item 3) and open the dosing glass.
Dosing
1
2
3
A0014128
23
Vacuum pump
1
Dosing tube
2
Allen screw
3
Air hose connection
1. Release the 2 mm (0.08 in) Allen screw with the key provided.
2. Set the sample volume by adjusting the dosing tube. Secure the dosing tube with the
screw.
3. Use the white scale (A) to dose without pressure and the blue scale (B) to dose with
pressure.