Diagnostics and troubleshooting
Liquistation CSF28
42
Hauser
No.
Message
Factory settings
Tests or remedial action
S
D
F
285
Update error
F
On
On
Firmware update failed
1.
Repeat.
2.
SD card error → use another card.
3.
Incorrect firmware → repeat with suitable
firmware.
4.
Contact Hauser Service.
302
Battery low
M
On
Off
Buffer battery of real time clock is low
The date and time are lost if the power is
interrupted.
‣
Contact Hauser Service (battery
replacement).
304
Module data
F
On
On
At least 1 module has incorrect configuration data
1.
Check the system information.
2.
Contact Hauser Service.
306
Software error
F
On
On
Internal firmware error
‣
Contact Hauser Service.
310
Temperature sensor
F
On
On
Temperature sensor PT1 in the climate control
module for sample compartment measurement is
defective
• No temperature regulation possible for the
sample compartment
• Unable to cancel the sampling program
‣
Contact Hauser Service.
313
Safety sensor
M
On
On
Safety switch LF2 for sample sensor active
• Contact electrodes for sample detection are
fouled
• The sample continues to be taken
1.
Clean sample detection sensor LF1 in the
dosing glass.
2.
Contact Hauser Service.
314
No sample flow
F
On
On
A vacuum cannot be generated in the peristaltic
pump.
1.
Check the pump tube for leaks.
2.
Immerse the suction line in the medium.
315
Refrigeration
F
On
On
• Sample compartment target temperature not
reached
• Cooling regulation not possible
1.
Check the sample compartment door.
2.
Perform the module test under
Menu
/
Diagnostics
/
System test
/
Cooling system
/
Check cooling
.
3.
Contact Hauser Service.
316
Heating
F
On
On
• Sample compartment target temperature not
reached
• Heating regulation not possible
1.
Check the sample compartment door.
2.
Perform the module test under
Menu
/
Diagnostics
/
System test
/
Cooling system
/
Check heating
.
3.
Contact Hauser Service.