13
Start-Up
Inserting the Batteries
With four AA batteries, the Liquiline To Go
Ex CYM291 has an operating time of
approx. 500 h.
Open the battery compartment on the
rear of the device. Be sure to observe the
correct polarity when inserting the batter‑
ies (see markings in the battery chamber).
Close the battery compartment cover and
screw it handtight.
Check the shipment for transport damage and completeness (see Package
Contents).
NOTICE!
Do not operate the device when one of the following conditions applies:
• the device shows visible damage
• the device fails to perform the intended function
• prolonged storage at temperatures above 70 °C
• severe transport stresses
In this case, a professional routine test must be performed.
This test should be carried out at our factory.
Precautions for application in hazardous locations
WARNING!
• Only open the battery compartment of the Liquiline To Go Ex CYM291
outside the hazardous location.
• Never try to open the device. If a repair should be required, return the
device to our factory.
• Never use the USB port within the hazardous location.