14
Start-Up
WARNING!
When operating the Liquiline To Go Ex CYM291 in a hazardous loca‑
tion, only the battery types listed below may be used. The batteries
must be from the same manufacturer and of identical type and
capacity. Never use new and used batteries together (“Certificates”,
Control Drawing 209.009‑150).
Batteries for Application in Hazardous Locations
Batteries (4x each)
Temp. class
Ambient temperature range
Duracell MN1500
T4
–10 °C ≤ Ta ≤ +40 °C
Energizer E91
T3
–10 °C ≤ Ta ≤ +50 °C
Power One 4106
T3
–10 °C ≤ Ta ≤ +50 °C
Panasonic Pro Power LR6
T3
–10 °C ≤ Ta ≤ +50 °C
A battery icon in the display indicates the battery power level:
Icon fully filled
Batteries at full capacity
Icon partially filled
Battery capacity is sufficient
Icon empty
Battery capacity not sufficient;
calibration is possible, no logging
Icon blinks
Only a few operating hours remaining,
measurement is still possible
NOTICE! It is absolutely necessary to replace
the batteries.