27
Oxy
Oxygen Calibration
Measurement
CAL
AIR CAL
Place sensor in air and wait until the measured
values have stabilized.
Measured value
Temperature
xxx rH blinks
Use
st
to adjust the correct value for relative
humidity.
xxx nA blinks
Hourglass blinks
Calibration will be performed.
CAL DATA
Date
SLOPE xxx nA
Slope calibration data is displayed.
Automatic return to measuring mode.
AIR CAL Calibration
(Calibrating the slope in air)
The calibration method is selected in the configuration menu.
Please note: To abort calibration, you can press
at any time.