17
pH
Configuration (pH)
Configuration is required to match the connected sensor and the
desired measurement performance. Furthermore, you can select the
suitable calibration method. The following table gives you an over-
view. Factory settings are shown in bold print.
Measurement
“Setup” display
Select using arrow keys, confirm by pressing .
Display 1
pH x.xx | pH x.xxx | mV
Display 2
OFF | date + time | date | time
CAL Timer
OFF | 1 ... 99 days
CAL
PROG.BUFFERS|Manual|DATA INPUT|
(ISFET-Zero)
|
FREE CAL
CAL POINTS
1 | 2 | 3 | 1-2-3 (for PROG.BUFFERS, Manual, FREE CAL)
BUFFER SET
(PROG.BUF-
FERS, FREE
CAL)
-01- Hauser
2.00|4.01|6.98|9.95|11.87
-02- Mettler-Toledo
2.00|4.01|7.00|9.21
-03- Knick CaliMat
2.00|4.00|7.00|9.00|12.00
-04- Ciba (94)
2.06|4.00|7.00|10.00
-05- NIST technical
1.68|4.00|7.00|10.01|12.46
-06- NIST standard
1.679|4.006|6.865|9.180
-07- HACH
4.01|7.00|10.01|12.00
-08- WTW techn. buffers 2.00|4.01|7.00|10.00
-09- Hamilton
2.00|4.01|7.00|10.01|12.00
-10- Reagecon
2.00|4.00|7.00|9.00|12.00
-11- DIN 19267
1.09|4.65|6.79|9.23|12.75
Auto OFF
OFF| 0.1h | 1h | 6h | 12h
Temp Unit
°C | °F
Time Format
24h | 12h
Date Format
dd.mm.yy | mm.dd.yy
Default
NO | YES (reset to factory settings)
Note: All data logger entries will be deleted.
This icon prompts you to select a menu item using the arrow keys –
the selection is confirmed by pressing .
Configuring