22
pH
Measurement
CAL
MANUAL
The number of calibration points has been
selected in the configuration menu.
CAL 1/2/3
PRESS CAL
pH display blinks
PRESS CAL
Take the temperature‑corrected pH value from
the buffer description and enter it using
st
.
mV display blinks
Depending on the number of calibration points,
the procedure described above for CAL 1/2/3 is
repeated.
mV value blinks until calibration is completed, then successive display of:
CAL DATA
1/2/3 CAL POINTS
ZERO POINT
SLOPE
Then the meter switches to measuring mode.
MANUAL Calibration
(Manual calibration)
The calibration method is selected in the configuration menu.
Please note: To abort calibration, you can press
at any time.
This will be confirmed by the display message “CAL ABORTED”.
Exception: When you have selected “CAL POINTS 1‑2‑3” and the first calibration step
has been completed, the calibration process cannot be stopped any more.
pH Calibration