pH
Oxy Cond
Data Logger
Operating Modes of the Data Logger (Logger Type)
Manual logging when logger is activated (SHOT)
In this mode, a measured value is recorded when the
key is pressed.
Measurement
Logger activated
The measured value is saved to the address of the last recorded value + 1
Manual logging when logger is deactivated
Measurement
Logger deactivated
Measured value is maintained
Proposed address blinks
(address of the last recorded
value + 1)
If desired: Select start address
using
st
.
Measured value is saved to the desired address (e.g., for overwriting an incorrect
measurement).
Example: Logger type “INT”
Data Logger
Interval (INT)
In this mode, the measured values are cyclically recorded.
Measured parameter
Time
= entry