Liquiline System CA80PH
Operation
Hauser
87
Standard 1
1 to 20 %
Factory setting
2 %
Reagent RB,
connection R1
1 ... 40 %
Factory setting
5 %
Reagent RK,
connection R2
1 ... 40 %
Factory setting
5 %
Diag. code727
Read only
Absorption curve recording
In automatic mode
Options
• Off
• On
Factory setting
Off
Off: Recording is only possible in the manual mode
On: Recording also takes place in the automatic mode
Curve
1 to 7
Factory setting
2
Select the absorption curve to be recorded.
Only one curve can be selected at any one time. The data
are saved in a logbook.
After power failure
Options
• Last mode
• Manual mode
Factory setting
Last mode
Setting specifying how the analyzer should behave after
a power failure, and when the power is back on.
Last mode
:
The analyzer remains in the mode that was last set.
Example: The automatic mode was set. The analyzer
returns to operation once initialization has been perfor-
med, any sample has been discarded and the system has
been cleaned with sample.
Manual mode
:
The analyzer switches to the manual mode and waits for
user action.
Path: Menu/Setup/Analyzer/Extended setup
Function
Options
Info