Liquiline System CA80PH
Diagnostics and troubleshooting
Hauser
163
11.13 Systemtest
11.14 Reset
Temperature
-50.0 to +250.0 °C
(-58.0 to 482.0 °F)
Factory setting
20.0 °C (68.0 °F)
Set the desired simulation value.
Diagnostics/Systemtest
Function
Options
Info
Analyzer
Sample collector
Is only displayed if a collecting vessel is provided.
Empty sample collector
The sample collecting vessel can be automatically
emptied in this menu.
Start
Stop
Photometer
Cleaning factor
Read only
Raw value
Read only
Temperature
Read only
Power supply
Read only
• Digital Supply 1: 1.2V
• Digital Supply 2: 3.3V
• Analog Supply: 12.5V
• Sensor Supply: 24V
• Temperature
Detailed list of power supply to instrument.
The actual values can vary without a malfunction
being present.
Diagnostics/Reset
Function
Options
Info
Device reset
Options
• OK
• ESC
Restart and keep all the settings
Factory default
Options
• OK
• ESC
Restart with factory settings
Settings that have not been saved are lost.
Diagnostics/Simulation
Function
Options
Info
Содержание Liquiline System CA80PH
Страница 2: ......
Страница 208: ...www addresses endress com ...