Liquiline System CA80PH
Maintenance
Hauser
171
12.4 Changing the hose
You need the following parts:
1. Rinse the system (see the "Rinsing the system" section).
There should not be any sample left in the sample collecting vessel.
2. Remove the bottles and the bottle tray.
3. Remove the cover of the carrier board.
4. Replace all the white C-Flex hoses with hoses of the same diameter and length. Attach hose
labels to the new hoses.
The drain hose D6 must be routed behind dispenser 7.
5. Secure the cover and the bottle tray back into the housing.
6. Connect the bottles to the appropriate covers and hoses.
7. Select "Menu/Operation/Maintenance/Bottle replacement/Bottle insertion/Bottle
selection".
8. Select all the bottles and press "OK" to confirm.
9. Select "Confirm bottles inserted".
10. Press the soft key "MODE" and select "Continue automatic mode" or "Start automatic mode".
Recommendation for the hose change interval:
• C-Flex, white
Every 12 months
• C-Flex hose, ID 1.6 mm
Part of the CAV800 maintenance kit
• C-Flex hose, ID 3.2 mm
• C-Flex hose, ID 16.4 mm
• Hose fitting
• 1 pair of gloves that are resistant to the reagents used