Technical data
Liquiline System CA80PH
198
Hauser
15.6 Power supply
15.6.1 Supply voltage
The analyzer is fitted with a connector with a cable length of 4.3 m (14.1 ft) . When
replacing the power cable, pay attention to national regulations concerning electrical
safety.
• 100 to 120 VAC / 200 to 240 VAC ± 10%
or
24 VDC ± 10%, 130 W
• 50 ± 1 or 60 ± 1.2 Hz
15.6.2 Fieldbus connection
Supply voltage: not applicable
15.6.3 Power consumption
130 VA + 660 VA per hose heating system, max. 1450 VA (version with cooling)
15.6.4 Fuse
5 x 20 mm 10 A/250 V fine-wire fuse for hose heating system
15.6.5 Cable entries
• 4 x bores for M16, G3/8, NPT3/8", Memosens connection
• 4 x bores for M20, G1/2, NPT1/2"
15.6.6 Hose entries
• 4 x bores for M32 for sample inflow and outflow
15.6.7 Cable specification
Cable glands mounted at the factory are tightened with 2 Nm.
Cable gland
Permitted cable diameter
M16x1.5 mm
4 to 8 mm (0.16 to 0.32")
M12x1.5 mm
2 to 5 mm (0.08 to 0.20")
M20x1.5 mm
6 to 12 mm (0.24 to 0.48")
NPT3/8"
4 to 8 mm (0.16 to 0.32")
G3/8
4 to 8 mm (0.16 to 0.32")
NPT1/2"
6 to 12 mm (0.24 to 0.48")
G1/2
7 to 12 mm (0.28 to 0.48")
Содержание Liquiline System CA80PH
Страница 2: ......
Страница 208: ...www addresses endress com ...