Liquiline System CA80PH
Commissioning
Hauser
59
9
Commissioning
9.1
Function check
WARNING
!
Incorrect connection, incorrect supply voltage
Safety risks for staff and incorrect operation of the device
‣
Check that all connections have been established correctly in accordance with the wiring
diagram.
‣
Make sure that the supply voltage matches the voltage indicated on the nameplate.
Before putting the device into operation:
1. Connect the medium-carrying hoses for sample supply.
2. If present, connect the communication cable and hose heater of the sample preparation
system to the analyzer.
Save configuration settings as a screenshot
Screenshots can be taken at any time using the local display and are stored on an SD card.
1.Insert an SD card into the SD card slot on the base module.
2. Press the navigator button for at least 3 seconds.
3. In the context menu, select the "Screenshot" option.
The current display is stored as a bitmap file on the SD card in the "Screenshots" folder.
9.2
Switching on the measuring device
‣
Switch on the supply voltage.
Wait for the initialization to complete.
9.3
Setting the operating language
‣
Press the soft key for "MENU". Select your language in the top menu item.