Commissioning
Proline Promag P 500 HART
124
Hauser
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Separator
A local display is provided.
Select decimal separator for
displaying numerical values.
• . (point)
• , (comma)
. (point)
Backlight
One of the following
conditions is met:
• Order code for "Display;
operation", option
F
"4-line,
illum.; touch control"
• Order code for "Display;
operation", option
G
"4-line,
illum.; touch control
+WLAN"
Switch the local display
backlight on and off.
• Disable
• Enable
Enable
*
Visibility depends on order options or device settings
10.5.5 Performing electrode cleaning
The
Electrode cleaning circuit
submenu contains parameters that must be configured for
the configuration of electrode cleaning.
The submenu is only available if the device was ordered with electrode cleaning.
Navigation
"Setup" menu → Advanced setup → Electrode cleaning circuit
‣
Electrode cleaning circuit
Electrode cleaning circuit
ECC duration
ECC recovery time
ECC cleaning cycle
ECC Polarity
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry / User
interface
Factory setting
Electrode cleaning circuit
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Enable the cyclic electrode
cleaning circuit.
• Off
• On
Off
ECC duration
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Enter the duration of electrode
cleaning in seconds.
0.01 to 30 s
2 s
ECC recovery time
For the following order code:
"Application package", option
EC
"ECC electrode cleaning"
Define recovery time after
electrode cleaning. During this
time the current output values
will be held at last valid value.
1 to 600 s
60 s