Commissioning
Proline Promag P 500 HART
106
Hauser
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Maximum frequency value
One of the following options is
selected in the
Assign current
output
parameter (→ 101):
• Volume flow
• Mass flow
• Corrected volume flow
• Flow velocity
• Conductivity
• Electronic temperature
Enter maximum frequency.
0.0 to 10
000.0 Hz
10
000.0 Hz
Measuring value at minimum
frequency
One of the following options is
selected in the
Assign current
output
parameter (→ 101):
• Volume flow
• Mass flow
• Corrected volume flow
• Flow velocity
• Conductivity
• Electronic temperature
Enter measured value for
minmum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter
Measuring value at maximum
frequency
One of the following options is
selected in the
Assign current
output
parameter (→ 101):
• Volume flow
• Mass flow
• Corrected volume flow
• Flow velocity
• Conductivity
• Electronic temperature
Enter measured value for
maximum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter
Failure mode
One of the following options is
selected in the
Assign current
output
• Volume flow
• Mass flow
• Corrected volume flow
• Flow velocity
• Conductivity
• Electronic temperature
Define output behavior in
alarm condition.
• Actual value
• Defined value
• 0 Hz
0 Hz
Failure frequency
One of the following options is
selected in the
Assign current
output
• Volume flow
• Mass flow
• Corrected volume flow
• Flow velocity
• Conductivity
• Electronic temperature
Enter frequency output value
in alarm condition.
0.0 to 12
500.0 Hz
0.0 Hz
Invert output signal
–
Invert the output signal.
• No
• Yes
No
*
Visibility depends on order options or device settings