Installation
CCS120
10
Hauser
3.3.3
Installing the sensor in immersion assembly CYA611
The immersion assembly CYA611-0B is designed for on-site installation of the sensor. Cover the
thread NPT 3/4" with a layer of Teflon tape. Screw the sensor into the assembly.
Note the instructions and safety guidelines in the operating instructions of the immersion
assembly.
3.4
Post-installation check
• Check the membrane for leak tightness und replace it if necessary.
• Is the sensor installed in an assembly and is not suspended from the cable?