Perform the following steps to fill the membrane cap with electrolyte:
1.
Open the electrolyte bottle and screw on the nozzle.
2.
Squeeze out excess air.
3.
Place the electrolyte bottle completely onto the membrane cap (Fig. 8).
4.
Slowly squeeze the electrolyte out of the bottle in one steady stream, while continously
retracting the bottle. The cap is completely full when the electrolyte reaches the lower
edge of the thread.
Now you have to install the membrane cap:
1.
Place the sensor shaft vertically onto the filled membrane cap.
2.
Do not cover the vent hole below the rubber seal with your fingers.
3.
Tighten on the membrane cap by hand as far as possible. There should not be a visible gap
between the sensor shaft and membrane cap. When you screw on the cap, excess
electrolyte will leak out through the hole below the rubber seal.
4.
Wipe away any electrolyte with a soft paper tissue or similar.
5.
Rinse the nozzle thoroughly with a clean, warm water jet so that no electrolyte remains
on the inside and outside of the nozzle.