Operation
CCS120
14
Hauser
5.3
Electrolyte fill in
#
Warning!
Do not swallow the electrolyte. Avoid contact of the electrolyte with skin and eyes. Otherwise
wash with a lot of water. In case of eye inflamation, contact a doctor.
"
Caution!
• Do not touch or damage the membrane or electrodes.
• The electrolyte is sensitive to oxidation: always keep the electrolyte bottle closed after use. Do
not transfer the electrolyte into other containers.
• The electrolyte should not be stored for more than 1 year and should not yellow (use by date,
see label).
• Avoid forming air bubbles as far as possible when pouring the electrolyte into the membrane
cap.
• The membrane cap must be used once only.
!
Note!
• Store the electrolyte bottle upside down in order that the viscous electrolyte can be poured
easily and bubble free into the membrane cap.
• Smaller bubbles will not interfere, larger bubbles leave the electrolyte by rising to the surface.
a0001688
Fig. 8: Membrane cap with electrolyte bottle
1
2
3
4
A
5
1
Nozzle
4
Membrane cap
2
Vent hole
5
Rubber seal
3
Electrolyte bottle
A
Electrolyte filling level