CCS120
Wiring
Hauser
11
4
Wiring
#
Warning!
• The electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
• Ensure that there is no voltage at the power cable before beginning the connection work.
4.1
Direct connection to the transmitter
Use the measuring cable CPK9-N*A1B (internal PM wire) to connect the sensor according to
the following connection table. See also the notes in the operating instructions of the transmitter.
* The white and the yellow cores are connected inside of the TOP68 coupling.
a0001689
Fig. 4: TOP68 plug pin arrangement (view from contact side)
Cable with TOP68 coupling
Liquisys
Pin
Assignment
Core
Terminal
1
TC-signal
coax-inner (white)
90
2
AGND
coax-outer (black)
12
3
4
+UB (15 V)
green
85
5
NTC1
yellow*
11
NTC1
white*
11
6
NTC2/AGND
brown
86
S
screen
S
S
1
2
3
5
6
4
1
2
3
5
6
4