Maintenance
Hauser
51
Removing the measuring unit
1.
Untighten the coupling and remove the pH and sodium electrode from the measuring
unit.
2.
Loosen the Allen screws (AF4) of the cover and then remove the cover.
3. Empty the measuring unit, e.g. with a pipette.
4.
Untighten the hose connections of the measuring unit. To do so, gently squeeze the
hoses in the direction of the connector, while pressing the locking ring, and remove
the hose from the connector.