Commissioning
CA76NA
24
Hauser
8.1.1
Connecting pipes that carry liquid
Flow diagram
S (1 ... 6)
O
O
O
1
x
P1
P2
P3
Na+
pH
DP
B
SL
LS
C
D
E
O
A0047930
9
Liquid control unit with measuring unit and supply vessel
S
Sample inlet, 1 to 6
O
Outlet
B
Overflow vessel for constant primary pressure
SL
Standard solution
C
Overflow level monitor
LS
Laboratory sample
D
Alkalization vessel
P1
Dosing pump
DP
Diisopropylamine (DIPA)
P2
Circuit pump
E
Supply vessel
P3
Alkalization pump
Connecting medium supply points
The analyzer can have up to 6 medium supply points depending on the device version.
Hose specifications (not included in scope of delivery):
• Externally toleranced, flexible PE or PTFE hose with outer diameter of 6 mm (0.24 in)
• Length at least 200 mm (7.87 in)