Maintenance
CA76NA
50
Hauser
11.2.2 Replacing the filter of the sample conditioning unit
Tools are not required for the following series of steps.
1.
Remove the medium supply hose with a quick-release coupling at the filter.
2.
Remove the filter cartridge.
3.
Insert a new filter cartridge, while paying attention to the flow direction (indicated
on the adhesive label on the filter)!
4. Fit the medium supply hose with the quick-release coupling back on.
11.2.3 Cleaning the measuring unit
1. If you have not already done so:
Switch off the automatic mode via:
Maintenance
/
Operating Mode
/
OFF
.
The device stops the program currently running. The analyzer is in the standby
mode.
2. Shut off the supply of medium at the control valve by turning the valve clockwise
Содержание CA76NA