CA76NA
Operation
Hauser
41
Group
name
Starting
address
Size
(bytes)
Format
Parameter name
Description
Unit
2
2
BYTE
Warnings
Assignment of warning
bits → 43
4
2
BYTE
Sample flow
Assignment of sample
flow→ 44
5
1
BYTE
Transmission
If current channel is in
"hold", value: 0
If current channel is
continuously
transferring
concentration data,
value: 1
6
4
REAL
pH value
pH value currently
measured
10
4
REAL
Sample temperature
Temperature of sample
flow that is currently
measured
°C
14
2
INT16
Time since last Na
calibration
Displays the time since
the last Na calibration
min
16
2
INT16
Time until next Na
calibration
Displays the time until
the next Na calibration
min
18
2
INT16
Time since last Na
regeneration
Displays the time since
the last Na regeneration
min
20
2
INT16
Time until next Na
regeneration
Displays the time until
the next Na regeneration
min
Status 2
22
4
REAL
S(Na)
Slope of the last Na
calibration
mV/dec
26
4
REAL
E0(Na)
E0 value from the last
Na calibration
mV
30
4
REAL
c0
c0 value from the last Na
calibration
ppb
34
4
REAL
T(Kal)
Average temperature
from the last Na
calibration
°C
38
4
REAL
S(pH)
Slope of the last pH
calibration
mV/dec
42
4
REAL
E0(pH)
E0 value from the last
pH calibration
mV
46
2
INT16
Calibration interval
Shows the currently
configured interval for
Na calibration
h
48
1
INT8
Number of
regenerations
Number of regenerations
performed between two
calibrations
49
1
INT8
Filling level of stock
solution
Filling level of the
sodium stock solution
%
Channel 1
50
4
REAL
Na concentration 1
Measured sodium
concentration of channel
1
ppb,
µg/l
54
1
BYTE
Status C1
Assignment of channel
status → 44
55
1
BYTE
Reserve C1
56
2
INT16
Measuring time C1
Set measuring time in
automatic mode
min