Technical data
CA76NA
66
Hauser
Protocol-specific data
Manufacturer ID
11
h
Device type
1571D
h
Device database files (GSD files)
www.endress.com/profibus
Device Integration Manager DIM
Output values
Status and measured values
Input variables
Remote control: measurement, calibration and regeneration of
the analysis function
Supported features
• PROFIBUS DP (DP-V0, cyclic data exchange), baud rate: 9.6
kbit/s – 12 Mbit/s
• PROFIBUS device address configured via onsite operation or
PROFIBUS Service "Set_Slave_Add"
• GSD
14.3 Power supply
Supply voltage
• 100 to 240 V AC (fuse must be replaced)
• 50 or 60 Hz
• Battery-free parameter backup
The analyzer is fitted with a fuse, T 1.25 A, for the 215 to 240 V AC voltage level. If
the analyzer is operated with 100 to 130 V AC, replace the fuse with the T 2.5 A fuse
supplied. The fuse is located in the cover of the electronics unit.
Power consumption
40 VA
14.4 Performance characteristics
Response time
CA76NA-**AD
0.1 to 2000 µg/l (ppb)
180 seconds (95 %) within a calibration interval of
72 hours
2001 to 9999 µg/l (ppb)
600 seconds (95 %) within a calibration interval of
72 hours
CA76NA-**AE
< 55 s
1)
Response time from sample inflow to display change, T
90
depending on the stages of the concentration
changes, 12 min. max.
Reference operating
conditions
Sample pH 7, 25 °C (77 °F), 1 bar (14.5 psi)
Maximum measured error
CA76NA-**AD
0.1 to 2000 µg/l (ppb)
2 % of measured value; ±2 µg/l (ppb) (under
reference conditions)
2001 to 9999 µg/l (ppb)
5 % of measured value; ±5 µg/l (ppb) (under
reference conditions)
CA76NA-**AE
0.1 to 40 µg/l (ppb)
2 µg/l (ppb)
> 40 µg/l (ppb)
5 % of measured value