CA76NA
Operation
Hauser
35
9.3
Maintenance
Maintenance
Function
Options
Info
Password W
Enter the password to access the submenu.
Factory password: 1111
1.
Enter the first digit of the password using
the and keys.
2.
Go to the next digit with the key.
3.
Once you have entered the password,
press to confirm.
4.
Press the key for longer to return to
the measured value display.
If the wrong password has been entered, the
Incorrect Password!
message appears on the
screen. The prompt to enter the password
continues to be displayed.
Operating Mode
You can start individual analysis programs and
the automatic program sequence. You can freeze
all the output values for maintenance.
Maintenance
Selection
• ON
• OFF
If
OFF
is selected, normal device operation is
guaranteed. If
ON
is selected, all the information
output by the device is frozen. Measured values,
errors and alarms are not forwarded. Switch on
this function if performing maintenance or tests.
Mode
Selection
• AUTOMATIC
• OFF
AUTOMATIC
Switches on the automatic program sequence for
the analyzer.
The analyzer immediately starts analyzing the
first channel that has been assigned a measuring
duration.
After this, the individual channels are analyzed
according to the set channel sequence and
measuring duration.
OFF
Switches off the automatic program sequence for
the analyzer.
Manual
Selection
• OFF
• CALIB.
•
Channel 1
... 6
• Grab-test
• Fill
• Regener.
•
OFF
: No program selected manually
•
CALIB.
: Multi-point calibration is performed
•
Channel 1 ... 6
: Channels 1 to 6 are analyzed
•
Grab-test
: Lab sample is analyzed
•
Fill
: Hose of the standard solution pump and
dosing loop are filled
•
Regener.
: Sodium electrode is regenerated