CA76NA
Electrical connection
Hauser
19
2. Open the cover of the electronics unit to the left.
6.2.2
Connecting the analog outputs, digital outputs and power
supply
Connecting the signal outputs
The measured value of the particular channel is available as a current signal on the analog
or digital output card. The analyzer can have up to 6 current outputs depending on the
device version.
1. Guide the cables through the cable entries at the bottom of the electronics unit.
Position and dimensions of the cable entries → 15.
2. Guide the cables through the cable glands to the electronics unit.
3. Connect the outputs as shown in the terminal connection diagram → 19.
Connecting the power supply
The analyzer is fitted with a fuse, T 1.25 A, for the 215 to 240 V AC voltage level. If
the analyzer is operated with 100 to 130 V AC, replace the fuse with the T 2.5 A fuse
supplied. The fuse is located in the cover of the electronics unit.
1. Guide the cables through the cable entries on the back of the electronics unit.
Position and dimensions of the cable entries → 15.
2. Connect with a 3-core cable to terminal strip X100 (L1/N/PE) in the electronics unit
according to terminal connection diagram → 19.
Terminal diagram without PROFIBUS
L1 N PE
COM2
NC2
NO2
NO1
COM1
NC1
X 100
A+
COM
COM
COM
COM
A+
A+
COM
COM
B+
B+
B+
X12
X15
X23
CH1
CH2
CH3
CH4
CH5
CH6
X3
X1
COM3
NC4
NO3
COM4
COM5
NC6
NO5
COM6
COM7
NC8
NO7
COM8
V+
IN2
IN1
V+
V+
IN4
IN3
V+
V+
IN6
IN5
V+
F01
X6
X9
X11
X13
X16
X17
A0033459
L1 N PE NO1 COM1 NC1 NO2 COM2 NC2 A
+
COM B
+
COM A
+
COM B
+
COM A
+
COM B
+
COM
X100
Power
supply
100 to
240 V AC,
50/60 Hz
X1
Relay 1
Alarm
X3
Relay 2
Warning
X12A
4 to 20
mA
Channel
1
X12B
4 to 20
mA
Channel
2
X15A
4 to 20
mA
Channel
3
X15B
4 to 20
mA
Channel
4
X23A
4 to 20
mA
Channel
5
X23B
4 to 20
mA
Channel
6
Содержание CA76NA
Страница 2: ......
Страница 71: ......