Installation and commissioning
multi EA 5000
32
5.4
Gas supply
The operator is responsible for the gas supply and the corresponding connections and
pressure reducers.
The connection hoses with outer diameter (AD) 6 mm and inner diameter (ID) 4 mm
are provided with the device. Their lengths are 2 m. If other lengths are preferred,
please contact the Analytik Jena AG customer service department. The required gases
and their qualities are listed in the technical data (→ "Technical data for the
multi EA 5000 "
5.5
Device layout and space requirements
The space required for the modular analysis system depends on all the components that
make up the measuring station. The analysis station always includes:
¡
Basic module
¡
1 sampling module (to the right of the basic device or on top of it)
¡
1 detection module (to the left of the basic device)
Several modules may also be placed in a row.
Component
Width x Height x Depth
[mm]
Mass [kg]
Arrangement
Basic module
multi EA 5000
510 x 470 x 550 mm
25 kg
Detection modules
CLD 5000
300 x 500 x 550 mm
13 kg
As last detector in a
row
UVFD 5000
MPO 5000
300 x 470 x 550 mm
13 kg
Left/right of other
detectors
Cl module
300 x 470 x 530 mm
12 kg
To the immediate left
of the basic module
Coulometric Sulfur
Module
300 x 470 x 530 mm
11 kg
Left of the Cl module
TC module
300 x 470 x 530 mm
12 kg
Left/right of other
detectors
TOC module
300 x 470 x 530 mm
12 kg
To the immediate left
of the basic module
or to the left of the
Cl module
Sampling modules
ABD
520 x 210 x 500 mm
12 kg
To the right of the
basic module
MBD
500 x 80 x 80 mm
0.35 kg
To the right of the
basic module
MMS 5000
ca. 510 x 280 x 500 mm ca. 9.5 kg
On the basic module
or on the ABD*