Maintenance and care
multi EA 5000
166
ð
Check the TIC reactor for deposits and cracks quarterly. Cleaning is only required if
TIC samples are no longer properly purged.
}
Undo the connection between the TIC reactor and the water traps.
}
Remove the 2 knurled screws on the cover of the cooling block. Remove the cover.
}
Remove the FAST connector with hose 80 from the side outlet of the TIC reactor.
}
Remove the waste hose (hose 86) to the condensate pump from the connection on
the bottom of the TIC reactor.
}
Remove the TIC reactor from the detection module and check it for deposits and
cracks.
}
Rinse the TIC reactor with ultrapure water.
}
Reinstall the TIC reactor in the detection module is reverse order.
}
Check the system for leaks.
ü
The detection module is ready for operation again.
1
2
3
4
5
6
7
Figure 78TOC detector, door opened
1 Water traps
2 Measuring gas hose from the basic
module (hose 80)
3 Condensate pump
4 TIC reactor
5 Cooling block (measuring gas dryer)
6 Halogen trap
7 Hose 81