multi EA 5000
Operation
39
Sample
Furnace installation po-
sition and sampling
Comment
AOX
Horizontal for
ABD
ABD with MMS
Column method:
Activated carbon in a
quartz container in a
quartz glass boat
Batch method:
Activated carbon with no
quartz container in a
quartz glass boat with
holding-down clamp
EC/OC
Horizontal for
ABD
ABD with MMS
Filter in a quartz glass
boat with hold-down
clamp
TC, TOC, NPOC in water
Vertical for
MMS 5000
or manual sampling
TIC in water
Vertical (manual sam-
pling)
Direct injection in the TIC
reactor
Gas, non-pressurized
Horizontal or vertical
for
GSS module
with ABD (horizontal)
For trace analysis: Verti-
cal + GSS module recom-
mended
Gas, pressurized
Horizontal or vertical
for
GSS/LPG combi module
GSS module with
GSS adapterbox
with ABD (horizontal)
LPG
Horizontal or vertical
GSS/LPG combi module
LPG module 2.0
with ABD (horizontal)
For trace analysis: Verti-
cal +
GSS/LPG combi module
recommended
6.3
Switching on the basic module and the modules
Before switching on the basic module, always check:
¡
The other components (detection modules, sampling modules, PC) are connected to
the basic module.
¡
The gas supply is connected in accordance with instructions. The inlet pressure is ex-
actly 600 kPa (6 bar).
¡
The samples have been prepared.
Switch on the basic module as follows:
}
Open the valves at the pressure reducers of the gas supply.
}
Switch on all required components (detection modules, sampling modules, PC).
}
Switch on the basic module via the power switch.